Simulation of loads created by a dental bridge prosthesis based on maxilla implants Full article
Journal |
Russian Journal of Biomechanics
ISSN: 1812-5123 |
||||||||
---|---|---|---|---|---|---|---|---|---|
Output data | Year: 2022, Number: 2, Pages: 46-54 Pages count : 9 DOI: 10.15593/RJBiomech/2022.2.05 | ||||||||
Tags | implant, dentistry, fixed prosthesis, edentulous maxilla, primary stability, Coulomb-Mohr strength criterion, finite element method. | ||||||||
Authors |
|
||||||||
Affiliations |
|
Abstract:
The study aims to assess the primary stability of the prosthetics on 8 implants for edentulous
maxilla. Cases of individual implants failure and different bone mineralization are considered.
The model of the maxilla was made in accordance with the three-dimensional tomographic image. Modeling of primary stability failure of implants under the influence of various loading cases
were carried out. Bone materials were assumed to be isotropic, but normal and critical bone
mineralization were considered. The problems were solved using the finite element method. The
Coulomb-Mohr strength criterion was used to assess the primary stability. Critical prosthetics
scenarios depending on the implants failure configuration and degree of bone mineralization are
indicated on the basis of criterion. The most critical schemes of implant failure in the distal region
for any type of mineralization with unilateral chewing load. The considered configuration of the
prosthesis on 8 implants increases the risk of damage the cortical bone with age, even when all
implants have survived.
Cite:
Fedorova N.V.
, Larichkin A.Y.
, Shevela A.A.
Simulation of loads created by a dental bridge prosthesis based on maxilla implants
Russian Journal of Biomechanics. 2022. N2. P.46-54. DOI: 10.15593/RJBiomech/2022.2.05 Scopus РИНЦ OpenAlex
Simulation of loads created by a dental bridge prosthesis based on maxilla implants
Russian Journal of Biomechanics. 2022. N2. P.46-54. DOI: 10.15593/RJBiomech/2022.2.05 Scopus РИНЦ OpenAlex
Original:
Федорова Н.В.
, Ларичкин А.Ю.
, Шевела А.А.
Моделирование нагрузок, создаваемых мостовидным зубным протезом с опорой на имплантаты верхней челюсти
Российский журнал биомеханики. 2022. №2. С.56-66. DOI: 10.15593/RZhBiomeh/2022.2.05 РИНЦ
Моделирование нагрузок, создаваемых мостовидным зубным протезом с опорой на имплантаты верхней челюсти
Российский журнал биомеханики. 2022. №2. С.56-66. DOI: 10.15593/RZhBiomeh/2022.2.05 РИНЦ
Identifiers:
Scopus: | 2-s2.0-85134992743 |
Elibrary: | 49189414 |
OpenAlex: | W4403963622 |
Citing:
DB | Citing |
---|---|
Scopus | 2 |